Over time, a “normal” drift may occur, making recalibration of the humidity probe necessary.
It is recommended to service your probes every year if you use an OAD, or every 2 years for other uses (probe maintenance is not covered by the warranty).
To organize the recalibration of a probe, please contact Weenat Support (or your distributor) to request a quotation and arrange the exchange of the probe for a recalibrated one (avoid leaving the sensor without a probe).
🔧 Required equipment
One calibrated Rotronic probe (available from Weenat Support or your distributor)
One 8 mm open-ended wrench
🎥 Video tutorial
This video shows the necessary steps to carry out this procedure. Since the video is in French, you can find a detailed step-by-step guide below.
📝 Step-by-step guide
Weenat sends you a calibrated probe;
Unscrew the 3 nuts located above the mounting plate by a few turns using the wrench;
- Turn the plate by one-tenth of a turn and remove it from the radiation shield cups;
- Unscrew the probe;
- Replace it with the new one;
⚠️ Make sure the gasket is correctly positioned under the plate !
- Reinstall the plate and the probe onto the weather shield cups;
Tighten the 3 nuts;
Place the old probe in the envelope, attach the return label that was sent to you, and mail it back.